Smart Harmful Gas Removal System
It is an improved equipment for safety environment that dryly purifies organic and inorganic acid gases harmful to the human body during production, experiments, and research in chemical and agrochemical laboratories at government research institutes, schools, and companies.Model | SMHGRS-S25 |
---|---|
External Dimensionp(W*D*H)(mm) | 400 * 400 * 955 |
Controller | 7’ touchscreen (temperature/humidity, VOCs, PM2.5, PM10, tires, smart functions, speed display) |
Speed | 5-speed |
Rated Voltage | AC220V, 50/60Hz |
Power Consumption | 220W |
Weight | 45kg |
Warranty | 1 Year |
COMPOSITION
FEATURES
EFFECTS
Quality standards for Harmful Gas Removal System
Product | Test items (unit) | Maximum value for application | Comparable specifications | Deliverable specifications | |||||
Standard criteria | |||||||||
Specifications | Applicable criteria | Specifications | Applicable criteria | Specifications | Applicable criteria | ||||
Harmful Gas Removal System(SMHGRS-S25) | Power Consumption(W) | 245.3 | Test Report (PC21-01269K) | - | - | Power consumption of 300 or less at rated input for 30 minutes | Company-presented standard test method | ||
Insulation Resistance(GΩ) | over 1.02 | - | over 1 | ||||||
Voltage Withstand(-) | Withstands test voltage / No insulation breakdown | - | Withstands test voltage with no insulation breakdown | ||||||
Film Adhesion(-) | No peeling | Test Report(CT21-133805K) | No peeling | KS M ISO 2409 | No peeling | KS M ISO 2409 | |||
Film Thickness(μm) | 85 | over 40 | KS M ISO 2808 | over 50 | KS M ISO 2808 | ||||
Normal operation after drop impact (-) | No abnormalities | Test Report (TBK-2020-000551) | - | - | No abnormalities in operation after impact | Company-presented standard test method | |||
Condition treatment test | Normal operation after high temperature (50℃) treatment | No abnormalities | - | - | No abnormalities in operation after 24 hours of conditioning | ||||
Normal operation after low temperature (-10℃) treatment (-) | No abnormalities | - | |||||||
Deodorization test(%) | Deodorization rate of harmful substances above 92.24 | Test Report (21-064773-01-1) | over 60 | KOUVA AS 02 | Hazardous substance removal efficiency of 80 or more | KOUVA AS 02 | |||
Lifespan Performance (L) (gas removal capacity) | Gas removal capacity above 8 | Test Report(PC21-01270K) | over 5 | SPS-KTC-B6870-6410 | over 5 (based on test air volume: 1 m3/min) | SPS-KTC-B6870-6410 | |||
Acid gas removal performance(%) | Acid gas removal rate above 95 | Test Report (20190707) | over 60 | Air Pollutant Emission Standards | Removal rate of over 80 (air pollutant emission standards) | Company-presented standard test method | |||
TVOC reduction capacity (%) | Initial TVOC concentration (mg/m3) | 0.6 | Test Report (TBK-2020-000551) | Below 1 ~ 0.4 | Indoor Air Quality Advisory Standards | Below 1 ~ 0.4 (indoor air quality advisory standards) | |||
TVOC concentration after spraying(mg/m3) | 3.07 | ||||||||
TVOC concentration after 30 minutes (mg/m3) | 0.2 |